PPC2+™ Operation and Maintenance Manual
©2001 DH Instruments, Inc.
Page 94
HS%
Purpose
Read or set the hold limit for automated pressure control in percent of range full
scale.
Command
Query
“HS%
n”
“HS%?”
Classic “HS%=
n
“HS%”
Arguments
n:
The hold limit in % FS of the current active range.
Remarks
The hold limit can be read and set as a pressure or as a percent of the range.
Example
(enhanced)
Cmd sent:
Query reply:
“HS .01”
“0.0100 %”
Example
(classic)
Sent:
Reply:
“HS=.01”
“0.0100 %”
Errors ERR#
6
The
‘n’ argument was invalid.
See Also
Sections 3.1.2.3, 3.1.2.4, 3.2.3, 3.2.6, “HS”
IF
Purpose
Increase the pressure quickly.
Command “IF
n”
Classic “IF=
n”
Arguments
n
‘0’ Closes the fast up valve.
‘1’ Opens the fast up valve.
Remarks
Opening the fast up valve will cause the pressure to increase quickly. Care must be
used, as the pressure will not stop increasing until the valve is closed, or the upper
limit is passed.
Example
(enhanced)
Cmd sent:
Reply:
“IF 1”
“1” (no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“IF=1”
“IF=1”
Errors ERR#
6:
The
n argument is a ‘0’ or a ‘1’.
See Also
Section 5.8.1.7, “UL”
IP
Purpose
Increase the pressure a given amount using the slow speed.
Command “IP
n”
Classic “IP=
n”
Arguments
n
The increase in pressure desired (current pressure units). This can
be from 0 to 2 % FS of the current RPT range.
Remarks
The slow speed will be used for a calculated amount of time (up to 5 seconds) to
create the desired change. The PPC2+ will not attempt to control the pressure to a
target, so the change in pressure will be approximate.
Example
(enhanced)
Cmd sent:
Reply:
“IP 2”
“2.000 kPa”(no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“IP=2”
“2.000 kPa”
Errors ERR#
6:
The
n argument is not within given limits.
See Also
None.
Содержание PPC2+
Страница 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Страница 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Страница 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Страница 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Страница 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Страница 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Страница 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Страница 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...