PPC2+™ Operation and Maintenance Manual
©2001 DH Instruments, Inc.
Page 48
3.2.9 ANALOG
m
PURPOSE
To select the current measurement range of the optional analog input function. This key
has no function if the analog input function is not present in the PPC2+.
m
PRINCIPLE
The PPC2+ is available with an analog input option that allows the user to measure up to
four external voltage or current sources. Four ranges of measurement are provided:
•
0 to 100 mV.
•
0 to 5 V.
•
0 to 10 V.
•
4 - 20 mA.
The selected range is effective on all of the inputs together. The measurement is
available on the PPC2+ front panel or remotely.
m
OPERATION
To access the ANALOG function, press the [ANALOG] function key from the main
run screen. If the PPC2+ is equipped with the analog input option, the display is:
Input range: 1:100 mV
2:5V 3:10V 4:4-20 mA
The cursor is on the number of the currently active measurement range. Make the
desired selection. The display returns to previous screen with the new analog
measurement range active.
>RO OVOM^\SMKV MYXXOM^SYX] ^Y ^RO KXKVYQ SXZ_^ YZ^SYX K\O WKNO K^ K MYXXOM^Y\ YX ^RO
::-\OK\ZKXOV]OO=OM^SYX!
Содержание PPC2+
Страница 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Страница 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Страница 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Страница 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Страница 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Страница 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Страница 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Страница 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...