7 . 1 D R I V E R S
The PPC2+ drivers option provides eight open collector drivers for operating external valves,
solenoids, indicators, etc. When operating from the setup-driver screen (see Section 3.3.4), pressing
enter will allow the operating mode of the drivers to be set. The two modes of operation are
Momentary and Toggle. A momentary driver will change state while the corresponding driver number
on the keyboard is being pressed. In toggle mode, the driver state will toggle each time the
corresponding key is pressed and released.
Each output can sink 500 mA at 12 V. However, the total output of all the activated drivers cannot
exceed one Amp. Therefore, if multiple drivers are being activated please refer to the following chart
as a guide.
Table 19. Driver Output
# OF ACTIVE
DRIVERS
MAX CURRENT
PER OUTPUT
1 500
mA
2 400
mA
3 275
mA
4 200
mA
5 160
mA
6 135
mA
7 120
mA
8 100
mA
The male connector (P/N 102478) for the DRIVERS port is supplied with the PPC2+ accessories if
the driver option was ordered.
Содержание PPC2+
Страница 1: ... 2001 DH Instruments Inc q 9ZO K SYX KXN 7KSX OXKXMO 7KX_KV ...
Страница 14: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 8 89 9 ...
Страница 24: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 8 89 9 ...
Страница 84: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 76 8 89 9 ...
Страница 132: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 124 8 89 9 ...
Страница 150: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 8 89 9 ...
Страница 156: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 148 8 89 9 ...
Страница 166: ...PPC2 Operation and Maintenance Manual 2001 DH Instruments Inc Page 158 8 89 9 ...