Installation
6-6
5.
Connect the Aeroquip fittings until a positive stop is felt. The external reservoirs are
identical and therefore, interchangeable. A dry Teflon spray lubricant is
recommended, See Figure 6-3.
Figure 6-5: Connecting the external reservoir’s flex line to the cold head
6.2 Compressor installation
CAUTION
Failure to follow these installation guidelines could result in voiding the warranty.
Prepare the helium compressor package location
Confirm that the physical space containing the helium compressor package packages has
an ambient temperature range of 45 °F to 100
F (7 °C to 38
C).
Place the helium compressor packages in a level position. For the helium compressor
package to operate under optimal conditions, it must be oriented within 5
of being level.
WARNING
The helium compressor package must be positioned to provide easy access to the front panel mounted
circuit breaker.
Locate the helium compressor package so that it may be easily repositioned to allow for
sufficient space change the adsorber. If the helium compressor package cannot be easily
repositioned, leave approximately two feet (0.6 m) clearance above, to the left and right
of each of them.
In addition to the above requirements, CP100 series air cooled models require two feet
(0.6 m) of clearance on each side of the unit to ensure proper air flow. CPA2800 series
air cooled models require two feet (0.6 m) of clearance around all four sides and three
feet (1 m) of clearance above for proper air flow.
Turn this wrench
Hold this wrench
Содержание CPA289C
Страница 1: ...CP PT410 CPA289C C Cryocooler PT410 CPA289C INSTALLATION and OPERATION MANUAL ...
Страница 4: ...1 100 D Section 1 Overview ...
Страница 7: ...2 100 C Section 2 Warranty ...
Страница 10: ...3 100 B Section 3 Safety ...
Страница 17: ...4 100 B Section 4 Inspection and Unpacking ...
Страница 20: ...05 PT410 CPA289C B Section 5 Specifications ...
Страница 31: ...Specifications 5 11 Figure 5 6 System I O wiring diagram ...
Страница 32: ...Specifications 5 12 Figure 5 7 System I O wiring diagram external supply ...
Страница 33: ...Specifications 5 13 5 4 Cold head drawing Figure 5 8 Cold Head Drawing ...
Страница 34: ...6 100 E Section 6 Installation ...
Страница 36: ...Installation 6 2 Inspect the cold head Figure 6 2 Outline drawing of standard cold heads A E D C B F ...
Страница 45: ...7 100 C Section 7 Operation ...
Страница 58: ...8 100 C Section 8 Routine Maintenance ...