Operation
7-9
7.7.6 Warning diagnostics on display screen
Warnings do not cause the helium compressor package to stop and do not prevent the helium compressor
package from starting. If more than one of the warnings below is present, only the highest priority one will be
displayed.
Warning Description
Explanation
"High Pressure Running High!"
High pressure is approaching high set point, if the warning does
not go away after the cold head cools down, verify that static
pressures are correct, and all lines are connected completely.
"High Pressure Running Low!"
High pressure is approaching the low set point; verify static
pressure, check historical data to determine if a leak may have
occurred.
“Low Pressure Running Low!”
Low pressure is approaching the low set point; verify static
pressure, check historical data to determine if a leak may have
occurred.
“Low Pressure Running High!"
Low pressure is approaching the high set point; Flex lines may
be reversed check orientation of lines, verify that the high
pressure is greater than the low pressure
"Delta Pressure Running High!"
Differential pressure approaching high set point; this condition
may occur on initial startup. Under normal operation it will
reduce as the cold head begins to cool down. If not check static
pressure and helium flex line connections
"Delta Pressure Running Low!"
Differential pressure approaching low error set point; helium
flex lines may be reversed
"Static Pressure Running High!"
Static pressure approaching high error set point; verify required
static pressure before starting the helium compressor package
"Static Pressure Running Low!"
Static pressure approaching low error set point; verify required
static pressure before starting the helium compressor package
"Water In Temp Running High"
Water inlet temperature is approaching high error set point;
reduce inlet water temperature or increase flow rate to prevent
overheating
"Water In Temp Running Low"
Water inlet temperature is approaching low error set point;
increase inlet water temperature or reduce flow rate to prevent
overcooling or overheating caused by reduced oil temperature
“Water Out Temp Running High!”
Water outlet temperature is approaching high error set point;
reduce inlet water temperature or increase flow rate to prevent
overheating
"Water Out Temp Running Low"
Water outlet temperature is approaching low error set point;
increase inlet water temperature or reduce flow rate to prevent
overcooling or overheating caused by reduced oil temperature
“Helium Temp Running Low!"
Helium discharge temperature running low; check cooling water
temperatures, increase ambient temperature if system has
been off and ambient temperature is below 45 °F
Содержание CPA289C
Страница 1: ...CP PT410 CPA289C C Cryocooler PT410 CPA289C INSTALLATION and OPERATION MANUAL ...
Страница 4: ...1 100 D Section 1 Overview ...
Страница 7: ...2 100 C Section 2 Warranty ...
Страница 10: ...3 100 B Section 3 Safety ...
Страница 17: ...4 100 B Section 4 Inspection and Unpacking ...
Страница 20: ...05 PT410 CPA289C B Section 5 Specifications ...
Страница 31: ...Specifications 5 11 Figure 5 6 System I O wiring diagram ...
Страница 32: ...Specifications 5 12 Figure 5 7 System I O wiring diagram external supply ...
Страница 33: ...Specifications 5 13 5 4 Cold head drawing Figure 5 8 Cold Head Drawing ...
Страница 34: ...6 100 E Section 6 Installation ...
Страница 36: ...Installation 6 2 Inspect the cold head Figure 6 2 Outline drawing of standard cold heads A E D C B F ...
Страница 45: ...7 100 C Section 7 Operation ...
Страница 58: ...8 100 C Section 8 Routine Maintenance ...