Safety
3-5
Section 7.1.2 Startup procedure
WARNING
The system operates in a closed loop cycle at pressures higher than 200 psig (12.8 bar) and at
temperatures below 120 K.
Section 7.6 Disassembling the system for transport or storage
WARNING
Do not touch the cold end of the cold head until it has warmed to room temperature. If there is frost on
the cold head, it is too cold to touch.
3.2.4 Section 8: Routine maintenance
Section 8.3 Cold head
CAUTION
The Pulse Tube style cold heads contain no user-serviceable parts. Attempting to disassemble the cold
head will void the warranty. Contact Cryomech, Inc if the cold head needs to be returned for servicing.
CAUTION
Service to Gifford-McMahon style cold heads must be performed using Cryomech, Inc parts and the
appropriate work instructions. Contact Cryomech, Inc. before servicing the cold head.
Section 8.4 Replace the adsorber
CAUTION
At no time should the Aeroquip couplings be removed from the adsorber. Replacement can be completed
without relieving system pressure, since the adsorber is equipped with Aeroquip couplings for sealed
removal.
WARNING
The input power must be disconnected from the helium compressor package before removing any helium
compressor package panels.
Section 8.5 Vent excess helium from the helium compressor package
CAUTION
Venting more than 5 psig (0.34 bar) of helium per minute will lead to improper oil migration within the
system. If this condition occurs, factory service will be required.
Содержание CPA289C
Страница 1: ...CP PT410 CPA289C C Cryocooler PT410 CPA289C INSTALLATION and OPERATION MANUAL ...
Страница 4: ...1 100 D Section 1 Overview ...
Страница 7: ...2 100 C Section 2 Warranty ...
Страница 10: ...3 100 B Section 3 Safety ...
Страница 17: ...4 100 B Section 4 Inspection and Unpacking ...
Страница 20: ...05 PT410 CPA289C B Section 5 Specifications ...
Страница 31: ...Specifications 5 11 Figure 5 6 System I O wiring diagram ...
Страница 32: ...Specifications 5 12 Figure 5 7 System I O wiring diagram external supply ...
Страница 33: ...Specifications 5 13 5 4 Cold head drawing Figure 5 8 Cold Head Drawing ...
Страница 34: ...6 100 E Section 6 Installation ...
Страница 36: ...Installation 6 2 Inspect the cold head Figure 6 2 Outline drawing of standard cold heads A E D C B F ...
Страница 45: ...7 100 C Section 7 Operation ...
Страница 58: ...8 100 C Section 8 Routine Maintenance ...