Specifications
5-4
High-Pressure Atmospheric Relief Valve
The helium compressor package high-pressure atmospheric relief valve is set at
420 ± 5 psig (29 ± .34 bar). At pressures above 420 psig (29 bar) the relief valve will open
automatically and relieve pressure to the atmosphere.
Internal Motor Overload Switch
An internal motor overload switch, located inside the helium compressor package module, protects
the system by sensing excessive current draw and temperature. This switch automatically resets
itself after the helium compressor package module cools to an acceptable level.
Cold Head High Pressure Relief Valve
The cold head high pressure atmospheric relief valve is set at 425 ± 5 psig (29.3 ± .34 bar). At
pressures above 425 psig (29.3 bar) the valve will open automatically and relieve pressure to the
atmosphere.
Содержание CPA289C
Страница 1: ...CP PT410 CPA289C C Cryocooler PT410 CPA289C INSTALLATION and OPERATION MANUAL ...
Страница 4: ...1 100 D Section 1 Overview ...
Страница 7: ...2 100 C Section 2 Warranty ...
Страница 10: ...3 100 B Section 3 Safety ...
Страница 17: ...4 100 B Section 4 Inspection and Unpacking ...
Страница 20: ...05 PT410 CPA289C B Section 5 Specifications ...
Страница 31: ...Specifications 5 11 Figure 5 6 System I O wiring diagram ...
Страница 32: ...Specifications 5 12 Figure 5 7 System I O wiring diagram external supply ...
Страница 33: ...Specifications 5 13 5 4 Cold head drawing Figure 5 8 Cold Head Drawing ...
Страница 34: ...6 100 E Section 6 Installation ...
Страница 36: ...Installation 6 2 Inspect the cold head Figure 6 2 Outline drawing of standard cold heads A E D C B F ...
Страница 45: ...7 100 C Section 7 Operation ...
Страница 58: ...8 100 C Section 8 Routine Maintenance ...