Operation
7-11
Error Message
Lock Out
Contributor?
Trip
Clear
“You have had 6 errors in less
than an hour: MM:SS”
NO
>6 errors (lock out
contributors) in <1
hour
30 minute delay,
manual restart
“3 Phase Power Improperly
Wired!”
NO
Phase sequence
incorrect
Phase sequence
correct
"Power Supply Error!"
NO
Power supply fault
Power supply non
fault
Warning Description
Trip / Clear
"High Pressure Running High!"
375 psig (25.9 bar)
"High Pressure Running Low!"
170 psig (11.7 bar)
“Low Pressure Running Low!”
50 psig (3.4 bar)
“Low Pressure Running High!"
240 psig (16.5 bar)
"Delta Pressure Running High!"
290 psig (20.0 bar)
"Delta Pressure Running Low!"
75 psig (5.2 bar)
"Static Pressure Running High!"
280 psig (19.3 bar)
"Static Pressure Running Low!"
170 psig (11.7 bar)
"Water In Temp Running High"
85 °F (29 °C)
"Water In Temp Running Low"
45 °F (7 °C)
“Water Out Temp Running High!” 110 °F (43 °C)
"Water Out Temp Running Low"
45 °F (7 °C)
“Helium Temp Running High!”
170 °F (77 °C)
“Helium Temp Running Low!"
45 °F (7 °C)
“Oil Temp Running High!”
120 °F (49 °C)
“Oil Temp Running Low!"
45 °F (7 °C)
Содержание CPA289C
Страница 1: ...CP PT410 CPA289C C Cryocooler PT410 CPA289C INSTALLATION and OPERATION MANUAL ...
Страница 4: ...1 100 D Section 1 Overview ...
Страница 7: ...2 100 C Section 2 Warranty ...
Страница 10: ...3 100 B Section 3 Safety ...
Страница 17: ...4 100 B Section 4 Inspection and Unpacking ...
Страница 20: ...05 PT410 CPA289C B Section 5 Specifications ...
Страница 31: ...Specifications 5 11 Figure 5 6 System I O wiring diagram ...
Страница 32: ...Specifications 5 12 Figure 5 7 System I O wiring diagram external supply ...
Страница 33: ...Specifications 5 13 5 4 Cold head drawing Figure 5 8 Cold Head Drawing ...
Страница 34: ...6 100 E Section 6 Installation ...
Страница 36: ...Installation 6 2 Inspect the cold head Figure 6 2 Outline drawing of standard cold heads A E D C B F ...
Страница 45: ...7 100 C Section 7 Operation ...
Страница 58: ...8 100 C Section 8 Routine Maintenance ...