Safety
3-4
Section 6.2.3 Connect the compressor package to the main power
CAUTION
The mains supply voltage to the helium compressor package must be within 10% of the nominal voltage
specification in Section 5. Operating the helium compressor package outside the voltage specification
will void the warranty.
WARNING
Follow all local electrical codes and guidelines when during installation.
WARNING
One lead of the helium compressor package is grounded. Never bypass this ground or attach the helium
compressor package to an ungrounded circuit. A dangerous electrical hazard will develop.
Section 6.2.4 Connect the helium flex lines to the cold head or remote motor assembly and the
compressor package
CAUTION
Follow this procedure carefully when connecting and disconnecting the helium flex lines. Failure to follow
the procedure will void the warranty.
CAUTION
Do not allow the flex lines to contact any corrosives or commercial cleaning agents. Flex line damage
caused by exposure to corrosives or commercial cleaning agents will not be covered by the warranty.
CAUTION
Applying extreme bend stress to a flex line will cause permanent damage. Bending a flex line to less than
a 12 in (30 cm) radius will void the warranty.
3.2.3 Section 7: Operation
Section 7.1.1 Checks before operating
CAUTION
A voltage deviation of more than 10% above or below the voltage rating can cause helium compressor
motor overheating and possible failure. Indications of voltage operation outside that range will void the
helium compressor warranty.
CAUTION
Operating the cold head in a magnetic field of greater than 500 gauss will void the warranty.
Содержание CPA289C
Страница 1: ...CP PT410 CPA289C C Cryocooler PT410 CPA289C INSTALLATION and OPERATION MANUAL ...
Страница 4: ...1 100 D Section 1 Overview ...
Страница 7: ...2 100 C Section 2 Warranty ...
Страница 10: ...3 100 B Section 3 Safety ...
Страница 17: ...4 100 B Section 4 Inspection and Unpacking ...
Страница 20: ...05 PT410 CPA289C B Section 5 Specifications ...
Страница 31: ...Specifications 5 11 Figure 5 6 System I O wiring diagram ...
Страница 32: ...Specifications 5 12 Figure 5 7 System I O wiring diagram external supply ...
Страница 33: ...Specifications 5 13 5 4 Cold head drawing Figure 5 8 Cold Head Drawing ...
Страница 34: ...6 100 E Section 6 Installation ...
Страница 36: ...Installation 6 2 Inspect the cold head Figure 6 2 Outline drawing of standard cold heads A E D C B F ...
Страница 45: ...7 100 C Section 7 Operation ...
Страница 58: ...8 100 C Section 8 Routine Maintenance ...