4.3.2 Setting Cylinder Gas Concentrations
For the selected cylinder profile the actual gas concentration of each gas defined within the cylinder
must be checked against the physical cylinder to be connected to I-Test and, if necessary, adjusted.
X
From the Inlet Setup screen use the arrow keys to highlight the cylinder concentration icon
À
.
X
Press the
OK
key. The Gas Concentration setting screen
Á
will be displayed.
i
If a multiple gas cylinder has been set for the selected inlet, a choice of gases will be
displayed. Use the selection process described on
to select the required gas
and then continue.
X
Use the
and
keys to highlight the required value to adjust
Â
X
Use the
and
keys to increase or decrease the value.
X
Repeat this to adjust all the relevant values and press
OK
when the adjustment is complete.
The Gas concentration will be set and the Inlet Setup screen will be redisplayed or, if a multiple
gas cylinder has been selected for the inlet, the choice of gases will be redisplayed.
X
If whilst the Gas Concentration setting screen is displayed the adjustment is to be aborted, press
the
key.
Figure 77: Gas Concentration setting screen