43
Select Instrument:
To enable any field configuration update to be applied to the gas monitor the
Apply Changes On
check box must be ticked. If this tick box is not ticked no
changes will be made to the gas monitor.
i
The ‘Allow configuration tick box in the ‘Actions’ page must also be
selected to perform a field configuration.
Apply Changes On:
This allows the user to determine the date from which it is required to update
the configuration on the gas monitor.
If a date in the future is chosen I-Test will continue to operate as normal until
that date, at that point I-Test will then update the configuration each time the
gas monitor is inserted.
Enter the required start date for the gas monitor configuration by changing
each value in the field or clicking the calendar button to the right of the field.
Confidence
Strategy:
Confidence strategy is how the gas monitor indicates to the user that it is
functioning normally.
If it is required to change the confidence strategy setting on the gas monitor
activate the Confidence Strategy tick box and from the drop down box select
one of the following options:
Off:
No indications.
Sounder:
The gas monitor sounder beeps at preset intervals.
LED:
The gas monitor LED’s flash at preset intervals.
S LED:
The gas monitor sounder beeps and the confidence LED’s
flash at preset intervals.
T4 Configuration has two further options
Positive Safety:
The gas monitor will activate positive safety (+ve safety)
functionality which is an easy indication of operating status of T4 indicated by a
front mounted LED. Please refer to the T4 User & Operator Manual for further
details.
Positive sounder:
The gas monitor will activate positive safety
functionality plus the addition of the sounder when the positive safety LED is
illuminated.
Interval:
This field sets the time interval in seconds for the Confidence Strategy
indications. Select the required interval from a minimum interval of 5 seconds
to a maximum interval of 60 seconds.