Calibration:
This option should be selected if I-Test is to be configured to undertake only a
calibration function, on the gas monitor inserted.
This will be indicated in I-Test Manager by the text ‘(Stand alone)’ appearing
after the Calibration option when selected.
Standalone calibration will calibrate all sensors in the gas monitor, this will
occur even if the sensor was not due a calibration.
Standalone calibration cannot be selected when a speedy bump is selected.
If standalone calibration is selected with smart bump this will change the
functionality of the calibration process as described below in Bump Test Fail
Calibration.
Standalone calibration can be selected with Allow configuration, Download
event log, and Download event log.
Bump Test Fail
Calibration:
This option should be selected if there is a requirement to calibrate a sensor,
should it fail a speedy bump or a smart bump test.
If Calibration and Speedy Bump or Smart Bump are selected together, I-Test
will not perform a standalone calibration on any sensor that fails the selected
bump test.
This will be indicated in I-Test Manager by the text ‘(Bump Test Fail)’ appearing
after the Calibration option when speedy bump or speedy bump or smart bump
is also selected.
Only sensors that fail the selected bump test will be calibrated, this will occur
even if the sensor was not due a calibration.
Bump Test Fail Calibration can be selected with Speedy Bump, Smart Bump,
Calibrate if Expired, Allow configuration, Download monitor event log, and
Download monitor data log.
Calibrate if Expired:
This option should be selected if it is required that a gas monitor whose
calibration has expired, when inserted into I-Test should be automatically
calibrated if an appropriate cylinder is attached. This will override a bump test
action if configured and only a calibration will be undertaken.