Action On Startup:
This field determines the functionality of the gas monitor when powered up
with regard to the built in ‘Clean air’ zero function.
If it is required to change the Action on Startup setting on the gas monitor
activate the Action on Startup tick box and from the drop down box select one
of the following options:
No action:
No ‘Clean air’ zero action at startup.
Confirm zero:
This setting allows the ‘Clean air’ zero to be carried out at the
user’s discretion.
Always zero:
This setting means a ‘Clean air’ zero will always be carried out
at startup.
Logging Interval:
This field determines the interval at which the gas monitor will record gas
readings to its internal log file.
If it is required to change the logging interval activate the logging interval
tick box and set the required time interval in seconds. Minimum interval 1 of
second and maximum interval of 32400 seconds between log readings.
Calibration Due
Increment:
If a gas monitor is configured to undertake either a Bump Test Fail Calibration
or a Stand alone Calibration, and the calibration is successful, the calibration
due date will be incremented within the gas monitor.
The increment interval (in days) is preconfigured within the gas monitor at
factory, typically 30 days.
It is possible to change this increment by uploading a new configuration file to
the gas monitor.
Activate the Calibration Due Increment tick box and set the required interval in
days (minimum 1, maximum 720) between calibrating the monitor
i
Once I-Test has been configured to apply a monitor field configuration file, it will continue
to do this to each and every monitor inserted into I-Test. If it is required to stop the
configuration update it is necessary to deselect the ‘Apply Changes On’ tick box and save
this configuration to I-Test.