When all the required changes have been made, select
Save Configuration
to save the changes
(see
).
i
It should be noted that not all the I-Test configuration data is uploaded by this process.
The following items will not be uploaded:
Banner &
Banner Sets:
The banner number (relating to the banner I-Test was displaying from the
installed banner set) will be uploaded, but the banner set will not be uploaded.
To determine the details of the specific banner from the banner set, review the
copy of the I-Test configuration that was stored to the database.
Inlet configuration:
Bump Test Due Increment and Smart bump tolerance will not be uploaded
from the I-Test.
To determine the details of the Bump Test Due Increment and Smart bump
tolerance review the copy of the I-Test configuration that was stored to the
database.
i
It should also be noted that a user has the ability to change a limited set of I-Test
configuration settings via the I-Test keypad in the field. Therefore the master copy of
the database configuration may not entirely match the configuration uploaded to I-Test
Manager from I-Test at a later date, as changes may have been made directly on I-Test in
the field (see
i
It is important to note that the cylinder definitions, created in Manage Cylinders (see
), are stored in I-Test Manager’s database, this is the master record for cylinder
definitions. When a configuration is uploaded from an I-Test, the cylinder assigned to an
inlet is cross referenced to the this database record to enable the correct cylinder to be
displayed on the inlet within I-Test Manager.
i
When a configuration is uploaded from an I-Test, the cylinder assigned to an inlet is cross
referenced to the this database record to enable the correct cylinder to be displayed on
the inlet within I-Test Manager.
i
For the correct cylinder to be displayed in I-Test Manager following an upload from I-Test,
the current database must same database from which the cylinder was originally defined
otherwise the cylinder displayed may be incorrect.