28
Save As:
Enables a new cylinder profile to be created.
Once saved the new cylinder profile will be stored to the database and will be
visible in the Defined Cylinders list.
The saved cylinder profile name will also be the name shown on the screen of
the I-Test when the configuration is transferred and the cylinder assigned to
the gas inlet.
Save:
Saves the alterations against the original cylinder profile (Edit profile only).
Cancel:
Aborts any changes.
At this stage of I-Test configuration the cylinder profile will contain the standard gas concentrations;
the actual level of gas in the cylinder is entered during ‘Inlet Configuration’ see
Once the cylinders have been defined in the Manage Cylinders section the I-Test can now be fully
configured as detailed in the create new configuration section.
Note: When calibrating O
2
sensors I-test uses the purge inlet and calibrates on fresh air.
The O
2
sensor will then be bump tested with the cylinder O
2
gas to ensure correct operation (note
this bump test will only be undertaken on Gas-Pro monitors configured with firmware versions 1V16
& 2V06 onwards).