Symptom
Cause
Action
Error icon
is
displayed.
Gas cylinder is exhausted and
must be replaced.
Once I-Test has detected an
exhausted cylinder it is important
to change the cylinder Lot number
(see
cylinder is replaced as otherwise
I-Test will continue show this error
message.
Error icon
is
displayed.
Gas cylinder expiry date has
been exceeded.
Replace the gas cylinder and
reconfigure the cylinder Lot number
and expiry date (see
on
Error icon
is
displayed.
A monitor with sensors/gases
not supported by I-Test has been
inserted.
Error icon
and I-Test
gas monitor pod LED
flashing red.
The monitor has not been
inserted correctly.
Remove the monitor, re-insert
correctly and repeat test.
Icon
shown during
entry of values via I-Test
front panel
Value entered is too high to be
accepted
Enter a lower value.
Icon
shown during
entry of values via I-Test
front panel
Value entered is too low to be
accepted
Enter a higher value.
Test result declared as
incomplete in calibration
or bump test report.
Monitor may have been removed
during test.
Gas cylinder may have exhausted
during test.