83
Calibration
Pass:
If this tick box is selected gas monitors that have passed either a standalone
calibration or a bump test fail calibration will appear in the exception report.
Fail:
If this tick box is selected gas monitors that have failed either a standalone
calibration or a bump test fail calibration will appear in the exception report.
Cal Due in
Select this tick box and enter a value in the days field and gas monitors that are due a calibration in
the specified number of days will appear in the exception report.
Alarm Reporting
i
For an instrument to appear in the exception report for the following settings the I-Test
Action log files, and gas monitor event log files must have been downloaded from I-Test
and imported into I-Test Manager.
Alarm 1
If this tick box is selected gas monitors that have been exposed to a gas level exceeding the alarm
level one threshold, will appear in the exception report.
TWA Alarm
If this tick box is selected gas monitors that have been exposed to a gas level exceeding the TWA
threshold will appear in the exception report.
Alarm 2
If this tick box is selected gas monitors that have been exposed to a gas level exceeding the alarm
level two threshold, will appear in the exception report.
Faults
If this tick box is selected gas monitors that have logged a fault will appear in the exception report.