Using OxiMax™ Pulse Oximetry Sensors and Accessories
88
Operator’s Manual
WARNING
Use only Nellcor-approved OxiMax pulse oximetry sensors and pulse oximetry cables
when connecting to the OxiMax sensor connector. Connecting any other cable or
sensor influences the accuracy of sensor data, which may lead to adverse results.
WARNING
Do not use a damaged OxiMax pulse oximetry sensor or pulse oximetry cable. Do not
use a sensor with exposed optical components.
WARNING
Do not attach any cable to the oximeter sensor port connector that is intended for
computer use.
WARNING
Tissue damage can be caused by incorrect application or duration of use of an SpO
2
OxiMax pulse oximetry sensor. Inspect the sensor site periodically as directed in the
sensor
Directions for Use
.
WARNING
Do not lift the pulse oximeter by the pulse oximetry cable or power cord because the
cable or cord could disconnect from the pulse oximeter, causing the pulse oximeter to
drop on the patient.
WARNING
Pulse oximetry readings and pulse signal can be affected by certain ambient
environmental conditions, sensor application errors, and certain patient conditions.
WARNING
Do not immerse or wet the OxiMax pulse oximetry sensor.
Caution
The sensor disconnect error message and associated alarm indicate that the sensor is either
disconnected or the wiring is faulty. Check the sensor connection and, if necessary, replace
the OxiMax pulse oximetry sensor, pulse oximetry cable, or both.
Caution
OxiMax adhesive pulse oximetry sensors are intended for single-patient use only. Do not
transfer a sensor containing sensor trend data from one patient to a second patient. Doing
so may result in data from the first patient being used to evaluate the second patient, since
recorded SpO
2
event history data cannot distinguish among events collected from multiple
patients.
Note:
Physiological conditions, medical procedures, or external agents that may interfere with the
pulse oximeter’s ability to detect and display measurements include dysfunctional
hemoglobin, arterial dyes, low perfusion, dark pigment, and externally applied coloring
agents such as nail polish, dye, or pigmented cream.
Contact Nellcor’s Technical Services Department at 1.800.635.5267 or your local
Nellcor representative for a Sensor Accuracy Grid listing all of the OxiMax pulse
oximetry sensors used with the OxiMax N-600x pulse oximeter.
Содержание Nellcor OxiMax N-600x
Страница 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Страница 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Страница 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Страница 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Страница 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Страница 145: ......