Low and Critical Battery Conditions
Operator’s Manual
109
Low and Critical Battery Conditions
Table 18.
Low and Critical Battery Conditions
State
Critical
Battery
Low
Battery
AC
Power
Operation
1
No
No
Yes
SpO
2
-normal
AC/Battery charge LED-on
LOW BATTERY LED-off
LOW BATTERY message-off
Audible alarm-off
Error code-none
Effect of ALARM SILENCE key-normal
Shutdown-N/A
2
No
No
No
SpO
2-
normal
AC/Battery charge LED-off
LOW BATTERY LED-off
LOW BATTERY message-off
Audible alarm-off
Error code-none
Effect of ALARM SILENCE key-normal
Shutdown- N/A
3
No
Yes
No
SpO
2
-normal
AC/Battery charge LED-off
LOW BATTERY LED-on
LOW BATTERY message-on
Audible alarm-low priority
Error code-logged
Effect of ALARM SILENCE key-First press silences audio alarm,
second press cancels LOW BATTERY message (LED) stays on until
Low Battery Condition is corrected.
Shutdown-Imminent
4
No
Yes
Yes
SpO
2
-normal
AC/Battery charge LED-on
LOW BATTERY LED-on
LOW BATTERY message-off
Audible alarm-off
Error code-logged
Effect of ALARM SILENCE key-N/A (LED stays on)
Shutdown-N/A
5
Not used
Содержание Nellcor OxiMax N-600x
Страница 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Страница 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Страница 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Страница 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Страница 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Страница 145: ......