Prompts and Error Messages
Operator’s Manual
105
DEFAULTS LOST
None
No
No
If the oximeter
detects that
power-on settings
have been lost.
Leave the factory default
settings, have a qualified
service technician
change institutional
defaults, or change
temporary limit settings.
DELETE TRENDS?
10
Yes
Yes
When attempting
to delete trend
data from
memory by
pressing the
DELETE softkey.
Respond to the prompt,
electing to delete trends
by selecting YES or to
retain by selecting NO.
HELP SPEAKER
FAILURE
None
No
No
After pressing the
HELP softkey
provided after a
Primary Speaker
Failure.
Notify a qualified service
technician. Once the
oximeter powers down, it
will not power on again
until the technician
replaces the failing
speaker.
INVALID SILENCE
DURATION
3
No
No
When attempting
to save current
settings as power-
on defaults, even
though the
ALARM SILENCE
duration is OFF.
If attempting to save
defaults, turn ALARM
SILENCE duration to a
value, rather than OFF.
INVALID %SpO
2
LIMIT
3
No
Yes
After attempting
to set the default
%SpO
2
lower
alarm limit below
85.
Set power-on %SpO
2
lower alarm limit default
at or above 85.
LOW BATTERY
None
No
Yes
When the
oximeter is on
battery power
and the battery
charge is low.
Connect to AC power to
fully charge or
temporarily acknowledge
by pressing the ALARM
SILENCE key.
PRIMARY SPEAKER
FAILURE
None
No
No
Fatal error caused
by a hardware
failure. Once this
error displays, the
only option
available is to
press the HELP
softkey, which
produces the
HELP SPEAKER
FAILURE message.
Press the HELP softkey to
review the HELP SPEAKER
FAILURE message. Notify
a qualified service
technician. Once the
oximeter powers down, it
will not power on again
until the technician
replaces the failing
speaker.
Table 16.
Common Prompts and Error Messages
Message
Time-out
in seconds
Exit on
Alarm
Exit on
Alarm
Silence
Displayed
Resolution
Содержание Nellcor OxiMax N-600x
Страница 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Страница 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Страница 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Страница 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Страница 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Страница 145: ......