Managing OxiMax™ Pulse Oximetry Sensor Data
Operator’s Manual
69
•
BANDAGE ASSEMBLY
•
NAIL POLISH
•
SENSOR TOO TIGHT?
•
REPOSITION SENSOR
•
ISOLATE INTERFERENCE SOURCE
•
CLEAN SENSOR SITE
To clear messages, review the listed corrective actions and take appropriate steps to
clear the condition, then press the EXIT softkey. Once closed, the sensor message
screen will not return until a new condition occurs.
Figure 58.
Corrective Action Screen
Enabling or Disabling the Sensor Message Feature
The OxiMax pulse oximetry sensor message setup display allows you to enable or
disable the sensor message feature. When disabled, neither the “SENSOR NOT
POSTING” nor the “SUGGESTED ACTION” messages display.
1. With the oximeter in the normal monitoring mode, press the SETUP softkey.
2. Press the SENSOR softkey.
3. Press the MSG softkey.
Figure 59.
Enable Sensor Message Screen
4. Press the ADJUST UP or ADJUST DOWN key below the ALARM SILENCE key on the
oximeter to toggle the ENABLE message.
5. Press the EXIT softkey.