Low and Critical Battery Conditions
Operator’s Manual
111
The Pulse Search Indicator is lit for
more than 10 seconds (before any
measurements are taken).
Check the OxiMax™ pulse oximetry sensor
Directions for Use
to confirm
appropriate usage and proper application. Check sensor and pulse oximetry
cable connections. Test the OxiMax pulse oximetry sensor on another patient
and/or try another OxiMax pulse oximetry sensor or pulse oximetry cable.
Perfusion may be too low for the oximeter to track the pulse. Check the patient.
Test the oximeter on someone else. Change the OxiMax pulse oximetry sensor
site. Try another type of OxiMax pulse oximetry sensor.
Interference may be preventing the oximeter from tracking the pulse. Keep the
patient still, if possible. Verify that the sensor is securely applied and replace it if
necessary. Change the sensor site. Electromagnetic interference may be
preventing the oximeter from tracking the pulse. Remove the source of
interference and/or try to stabilize the environment.
Use a type of OxiMax pulse oximetry sensor that tolerates more patient
movement; for example, an OxiMax adhesive sensor.
The OxiMax pulse oximetry sensor may be too tight, there may be excessive
ambient light, or the sensor may be on an extremity with a blood pressure cuff,
arterial catheter, or intravascular line. Reposition the sensor as necessary.
The Pulse Search Indicator illuminates
after successful measurements have
been made.
Check the status of your patient.
Perfusion may be too low for the oximeter to track the pulse. Test the oximeter on
another patient. Change the OxiMax pulse oximetry sensor site and/or try
another type of OxiMax pulse oximetry sensor.
Interference may be preventing the oximeter from tracking the pulse. Verify the
OxiMax pulse oximetry sensor is securely applied and replace it if necessary.
Change the sensor site. Use a type of OxiMax pulse oximetry sensor that tolerates
more patient movement; for example, an OxiMax adhesive sensor.
Electromagnetic interference may be preventing the oximeter from tracking the
pulse. Remove the source of interference and/or try to stabilize the environment.
The OxiMax pulse oximetry sensor may be too tight, there may be excessive
ambient light, or the sensor may be on an extremity with a blood pressure cuff,
arterial catheter, or intravascular line. Reposition the sensor as necessary.
Error Code: “EEE XXX” followed by a
number appears.
Press the ON/STANDBY key and allow the oximeter to shut off completely. Then
press the key again to turn back on the oximeter.
If the error code persists, record the number and provide this information to a
qualified service technician, or your local Nellcor representative.
Error Code “EEE 529 or 729" displays when the battery discharges to a critically
low level.
Ensure the SUPPLY VOLTAGE SELECTOR switch on the rear panel is set to the
proper voltage based on your location.
Press the ON/STANDBY key and allow the oximeter to shut off completely. Allow
the battery to charge for about ten minutes and then turn back on the unit back.
If the error code is still present, turn the unit off and allow it to continue to charge.
If the oximeter has been charged for 30 minutes and the error code persists,
contact Nellcor’s Technical Services Department at 1.800.635.5267 or your local
Nellcor representative.
Table 19.
Common Problems and Resolutions
Problem
Resolution
Содержание Nellcor OxiMax N-600x
Страница 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Страница 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Страница 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Страница 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Страница 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Страница 145: ......