Operating the Oximeter
42
Operator’s Manual
3.
EXIT softkey
—Exit to the main menu or press BACK until you reach an EXIT menu option.
Main Menu
The main menu softkey options provide access to several submenus.
Figure 29.
Main Menu Options
1.
LIMITS
Menu
—Select standard oximeter upper and lower SpO
2
or pulse rate limits and
alarm management settings for either adults or neonates.
2.
TREND
Menu
—Select method of viewing oximeter trend and sensor data.
Note:
In the case of a SatSeconds or an SPD event, the oximeter proceeds straight to the MONITR
menu. Clear any alarm condition(s) to access the TREND menu.
3.
SETUP
Menu
—Use this softkey to control oximeter viewing options, sensor setup, clock
setup and language viewing options. Also configure the communication port, the nurse
call feature, and control analog voltage.
4.
LIGHT
Menu
—Leave the display backlight on, adjust brightness, or choose to turn the
display backlight off.
LIMITS Menu
Figure 30.
Limit Menu Options for Neonates
1.
SELECT softkey
—Use this softkey only after you have chosen the NEO screen or the
ADULT screen using the appropriate softkey. Then, use the SELECT softkey to scroll
through each limit setting option until you reach the limit value you wish to change.
2.
NEO softkey
—Use this softkey to set upper and lower limits for neonates. Scroll
through each limit setting option using the SELECT softkey until you reach the
limit value you wish to change. Change the value by pressing the ADJUST UP or
ADJUST DOWN key below the ALARM SILENCE key on the oximeter panel until
you reach the desired value. This setting only holds until power recycle. Decimal
points to the right of the displayed values indicate those limits have been
Содержание Nellcor OxiMax N-600x
Страница 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Страница 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...