Product Specifications
130
Operator’s Manual
Table 26.
Patient Applied and Patient Isolation Risk Current
Patient Applied Risk Current
Condition
AC Line Polarity
Neutral Line
Power Line Ground
Cable
IEC 60601-1
UL 60601-1
Normal
Normal
Closed
Closed
100
µA
Single Fault
Open
Closed
500
µA
Closed
Open
Normal
Reversed
Closed
Closed
100
µA
Single Fault
Open
Closed
500
µA
Closed
Open
Patient Isolation Risk Current
Condition
AC Line Polarity
Neutral Line
Power Line Ground
Cable
IEC 60601-1
UL 60601-1
Single Fault
Normal
Closed
Closed
5000
µA
Reversed
Closed
Closed