Managing Oximetry Data
78
Operator’s Manual
•
RS-422 Format
—Pins 1, 4, 9, and 12 provide RS-422 format data. When building an
RS-422 cable, add a resistor (120 ohms, 1/2 watt, 5%) between pin 1 and pin 9 of the cable
and keep cable length to a maximum of 4,000 feet. Plug the end of the cable with the
resistor added into the oximeter.
•
Differential Data Transmission
—Pins 3 and 4 (TxD+ and TxD-) are the differential
transmit data pair.
•
Differential Data Reception
—Pins 1 and 2 (RxD+ and RxD-) are the differential receive
data pair.
Figure 71 illustrates the pin layouts as viewed from the rear panel. The conductive
shell connects to earth ground when connected to a PC or printer.
Figure 71.
Data Port Pin Layout
Table 9.
Data Port Signal Pinouts
Pin
Number
Signal Name
1
RxD+ (RS-422 [+] input)
2
RxD_232 (RS-232 input)
3
TxD_(RS-232 output)
4
TxD+ (RS-422 [+] output)
5
Signal Ground (isolated from Earth Ground)
6
AN_SpO2 (analog saturation output)
7
NC_NO (relay closure nurse call, normally open)
8
NC_NC (relay closure nurse call, normally closed)
9
RxD- (RS_422 [-] input)
10
Signal Ground (isolated from Earth Ground)
11
Nurse Call (RS-232-level-output)
12
TxD- (RS-422 [-] output)
13
AN_PULSE (analog pulse rate output)
14
AN_PLETH (analog pleth waveform output)
15
NC_COM (relay closure nurse call, common lead)
Содержание Nellcor OxiMax N-600x
Страница 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Страница 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Страница 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Страница 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Страница 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Страница 145: ......