Operating the Oximeter
40
Operator’s Manual
Managing Audible Alarms
To set the alarm silence duration
The Alarm Silence Duration display enables you to adjust the alarm silence duration.
1. With the oximeter in the normal monitoring mode, press the ALARM SILENCE key until
the alarm silence duration setting displays. The alarm silence durations available are 30-,
60-, 90-, and 120-second intervals.
Figure 27.
Alarm Silence Duration Screen
2. Press and hold the ALARM SILENCE key and the ADJUST UP key below the ALARM
SILENCE key on the oximeter panel to increase the alarm silence duration setting.
3. Press and hold the ALARM SILENCE key and the ADJUST DOWN key to decrease the alarm
silence duration setting.
Note:
Releasing the ADJUST UP or ADJUST DOWN key sets the alarm silence duration.
Note:
Default alarm silence duration is 60 seconds. Select the appropriate value for the situation.
The alarm silence duration adjusts up to 120 seconds and down to 30 seconds.
To disable audible alarms
WARNING
Do not disable the audible alarm function or decrease the audible alarm volume if the
patient’s safety could be compromised.
WARNING
Pressing ALARM SILENCE will keep ALL alarms from sounding for the alarm silence
duration period.
Note:
The ability to set the alarm silence duration to OFF can be enabled or disabled by qualified
service personnel as described in the
N-600x Service Manual
.
1. With the oximeter in the normal monitoring mode, press the ALARM SILENCE key until
the alarm silence duration setting displays.
2. While pressing the ALARM SILENCE key, press and hold the ADJUST UP key until OFF
displays. Release both keys.
Содержание Nellcor OxiMax N-600x
Страница 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Страница 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Страница 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Страница 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Страница 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Страница 145: ......