Safety Tests
Operator’s Manual
129
Leakage Current
The following tables display the maximum earth and enclosure leakage current
allowed, as well as patient leakage.
Table 25.
Earth and Enclosure Leakage Current Specifications
Earth Leakage Current
Condition
AC Polarity
Line Cord
Neutral
Line Cord
IEC 60601-1
UL 60601-1
Normal
Normal
Closed
Closed
500
µA
300
µA
Single Fault
Open
Closed
1000
µA
Closed
Open
Normal
Reversed
Closed
Closed
500
µA
300
µA
Single Fault
Open
Closed
1000
µA
Closed
Open
Enclosure Leakage Current
Condition
AC Line Polarity
Neutral
Line Cord
Power Line
Ground
IEC 60601-1
UL 60601-1
Normal
Normal
Closed
Closed
100
µA
Single Fault
Open
Closed
500
µA
Closed
Open
Normal
Reversed
Closed
Closed
100
µA
Single Fault
Open
Closed
500
µA
Closed
Open