Managing the View Display
Operator’s Manual
35
•
SPD feature
—Use the OxiMax SPD™ Alert (SPD) feature to detect patterns of
desaturation in the SpO
2
trend in adults. With the SPD feature enabled, the display view
includes both the SatSeconds circle icon and SPD triangle icon and their settings. The
SPD alarm sensitivity value appears just below the SPD icon. When the SPD feature is
enabled, the triangle icon fills from the bottom to the top as desaturation patterns
develop. The triangle icon empties from the top to the bottom as patterns dissipate. If the
SPD icon reaches full, it indicates a threshold violation and the SPD alarm sounds. The
oximeter will sound the SPD alarm sooner if the SPD alarm sensitivity is set to the default
value of one (1). A less sensitive setting will result in less frequent alarms. Visible when the
SPD feature is activated and the SPD feature is enabled. For more information about
activation, contact Nellcor’s Technical Services Department at 1.800.635.5267 or your
local Nellcor representative. For more information about using the feature, reference
Using the OxiMax SPD™ Alert Feature
When the oximeter is not attached to AC power, it runs on an internal battery. When
using the internal battery, a battery fuel gauge appears in the upper left corner of
the display.
To select the Pleth Display
1. With the oximeter in the normal monitoring mode, press the SETUP softkey.
2. Press the VIEW softkey.
3. Press the PLETH softkey.
Figure 20.
Pleth Display View
Using the Blip Display
Blip Display Overview
Use this display view for visual monitoring information in blip bar form. The blip
display includes a pulse amplitude blip bar, current measured SpO
2
and pulse rate,
current upper and lower SpO
2
and pulse rate limits. It also includes a battery fuel
gauge if running on battery power. If enabled, SatSeconds and SPD icons display.
•
SatSeconds feature
—For mild or brief SpO
2
limit violations, use the SatSeconds feature
to reduce nuisance alarms. With the SatSeconds feature enabled, the SatSeconds circle
icon fills in the clockwise direction as the SatSeconds alarm management system detects
SpO
2
readings outside of the limit setting. The SatSeconds circle icon empties in
counterclockwise direction when SpO
2
readings are within limits. When the icon reaches
full, a medium priority alarm sounds. For more information about using the feature,
reference
Using the SatSeconds™ Alarm Management Feature
•
SPD feature
—Use the OxiMax SPD™ Alert (SPD) feature to detect patterns of
desaturation in the SpO
2
trend in adults. With the SPD feature enabled, the display view
Содержание Nellcor OxiMax N-600x
Страница 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Страница 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Страница 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Страница 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Страница 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Страница 145: ......