NCALL Menu
—From the SETUP menu, press the NEXT softkey twice so the NCALL (Nurse
Call) menu option is available. Set voltage from +5 VDC to +12 VDC using Norm + or set
voltage from -5 VDC to -12 VDC using Norm- when there is no audible alarm. Voltages
switch polarity when the audible alarm sounds. Return to the COMM/NCALL menu using
the BACK softkey. Exit to the Main Menu by pressing EXIT.
7.
ANALOG Menu
—From the SETUP menu, press the NEXT softkey three times to reach the
ANALOG menu option. Select 0 Volt, 1 Volt, or Step options to calibrate analog signals.
Return to the ANALOG/MODE menu using the BACK softkey. Exit to the Main Menu by
pressing EXIT.
8.
MODE softkey
—The response mode establishes the frequency the oximeter uses to
calculate, record, and display SpO
2
saturation levels, but does not affect the calculation of
pulse rate. The response mode, however, may impact the SPD alarm behavior. From the
SETUP menu, press the NEXT softkey twice to reach the MODE menu option.
a.
Normal Mode
—Default response mode responds to changes in blood oxygen
saturation in five- to seven-seconds when calculating %SpO
2
. When in Normal Mode,
the screen does not display the Fast Mode icon.
b.
Fast Response Mode
—Fast mode responds to changes in blood oxygen saturation
levels in two- to four-seconds when calculating %SpO
2
. This can be particularly
helpful for situations that require close monitoring. The Fast Mode icon in italics
appears in the lower right corner of the screen when in Fast Mode.
Note:
The response mode display screen includes the current SpO
2
response mode setting and the
current measured SpO
2
and pulse rate. If SatSeconds is enabled, the SatSeconds indicator
displays. If SPD is enabled, both SatSeconds and SPD icons will display.
To set the response mode
a. With the oximeter in the normal monitoring mode, press the SETUP softkey.
b. Press the NEXT softkey three times.
c. Press the MODE softkey.
Figure 33.
%SpO
2
Response Mode Selection Screen
Содержание Nellcor OxiMax N-600x
Страница 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Страница 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Страница 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Страница 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Страница 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Страница 145: ......