F9 Fetal & Maternal Monitor • User Manual
- 32 -
5.3 Understanding the Recorder Paper Printout
CAUTION
:
1. If there is any discrepancy between the display and the printout, the printout should have priority.
2. If the data is ambiguous, clinicians should make diagnoses based on the clinical assessment.
Figure 5-1 below is an example of the recorder paper with traces. Comparing it with the Monitor
screen, you can find this extra information on the printed recording:
Paper vendor
– “F9-G” indicates the paper vendor is GE. “F9-P” indicates the paper vendor is Philips.
Paper Style
– The FHR pane range 30 ~ 240 bpm indicates the paper style is American Standard. The
FHR pane range 50 ~ 210 bpm indicates the paper style is International Standard.
Self
-
Check Trace
– The F9 Monitor prints a self-check trace after being switched on. It is used to
check if the recorder paper is properly loaded.
FHR1 Mark
– The thick FHR trace marked with “FHR1” indicates this trace is FHR1 trace.
FHR2 Mark
– The thin FHR trace marked with “FHR2” indicates this trace is FHR2 trace.
Numeric List
– A list of the current date, time, print speed, ID and FHR2 offset is printed at the start
of a monitoring and every ten minutes afterwards.
Page Mark
– Each recorder paper pack has 150 pages. When you notice the page mark comes to the
end, remember to load new paper in time.