F9 Fetal & Maternal Monitor • User Manual
- 64 -
9.2.2 Cleaning of the Accessories
Cleaning of Transducers and ECG Leads
To clean the US transducer, TOCO transducer, IUP Probe and DECG Lead, follow these steps:
1)
Wipe them with a soft cloth dampened in cleaning solution;
2)
Clean them with a soft cloth dampened in water;
3)
Air-dry them or wipe off the remaining moisture with a soft dry cloth.
The recommended cleansers for accessories are listed below:
Accessory Cleansers
Ultrasound Transducer
TOCO Transducer
BURATON LIQUID
MIKROZID
ETHANOL 70%
SPORACIDIN
CIDEX
DECG Cable
Mild alcohol-free soapy water
IUP Cable
Mild alcohol-free soapy water
CAUTION
:
1) Be sure the temperature of the cleaning solution does not exceed 45
º
C
(
113
º
F)
.
2) Do not immerse them in any liquid.
3) Only clean the outer surface of the connectors, make sure no liquid goes into the
connector.
4) After cleaning, confirm that no remaining cleanser is on the surface.
Cleaning the Belt
Wash soiled belts with soap and water. The water temperature must not exceed 60
ºC (140 ºF).
9.3 Disinfecting
Clean the equipment before disinfecting.
The table below lists the allowed disinfectant bases:
Type Base
Instrument Disinfectant
Glutaraldehyde up to 3.6%
Surface Disinfectant
Ethanol
1- and 2- Propanol