F9 Fetal & Maternal Monitor • User Manual
- 58 -
7.10.9 Fetal Monitoring Alarm Messages
This table lists the alarm information that might appear during fetal monitoring, their respective
causes and countermeasures.
Alarm Message
Cause
Countermeasure
Patient Alarm (Medium Level)
**FHR1 HIGH or
** FHR1 xxx > yyy,
**FHR2 HIGH or
** FHR2 xxx > yyy
FHR1 or FHR2 measuring result
(xxx) is higher than the set
upper limit (yyy) over the alarm
delay time.
**FHR1 LOW or
** FHR1 xxx < yyy,
**FHR2 LOW or
** FHR2 xxx < yyy
FHR1 or FHR2 measuring result
(xxx) is lower than the set lower
limit (yyy) over the alarm delay
time.
Check if the alarm limits are
suitable; check the patient’s
condition.
Technical Alarm (Low Level)
US1 UNPLUGGED
or
US2 UNPLUGGED
US transducer 1 or US
transducer 2 is not securely
connected.
Check the connection of the
transducer.
US1 SIGNAL LOSS
or
US2 SIGNAL LOSS
FHR1 or FHR2 signal is too
weak for the system to analyze.
Check if the US transducer is
positioned at the fetal heart; check if
the alarm limits are suitable; check
the patient’s condition.
TOCO
UNPLUGGED
TOCO transducer is not
securely connected.
Check the connection of the
transducer.
Fetus EQUIP MALF
The monitor board can not
communicate with the system
successfully.
Restart the F9 Monitor and try
again, contact CooperSurgical if the
connection still fails.
Battery Low
The battery power is too low to
support continued working of
the monitor.
Connect the F9 Monitor to AC
power supply.
Check Paper
There is no paper in the paper
drawer or the drawer is open.
Load paper and/ or close the drawer.