F9 Fetal & Maternal Monitor • User Manual
- 47 -
2)
With the catheter connected to the IUP cable, momentarily press the re-zero button on the
pressure cable (See Figure 7-2). The green light on the cable will flash for five seconds.
Figure 7-2: Rezeroing the system
3)
During that time, adjust the F9 Monitor to zero by pressing the
AUTO ZERO
button.
WARNING
:
•
Before insertion, the placental position should be confirmed, amniotic
membranes are adequately ruptured and sufficient cervical dilatation is assured.
•
Try to insert the catheter opposite the placental site. Do not insert the introducer
beyond the cervical OS. Use with caution when a uterine infection is present.
•
If resistance is met at any time during insertion, withdraw the catheter slightly
and try at a different angle. Forced insertion may result in patient’s discomfort
or injury.
CAUTION
:
•
Since procedures vary according to hospital needs/ preferences, it is the
responsibility of the hospital staff to determine exact policies and procedures
for both monitoring and amnioinfusion. The safe and effective use of the IUPC
depends on the skill of the clinician who applies /uses it.
•
Read the manufactures
Directions For Use of the IUPC prior to insertion.
7.6.3 IUP Monitoring Procedure
1)
Insert the IUPC according to the manufactures instructions for use.
2)
Connect the IUPC to the IUP cable. (See Figure 7-3)
Figure 7-3: Connect catheter to pressure cable