F9 Fetal & Maternal Monitor • User Manual
- 72 -
Appendix 2 Signal Input/Output Connector
Accessory equipment connected to these interfaces must be certified according to the respective
IEC/EN standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for
medical equipment). Furthermore all configurations shall comply with the valid version of the
system standard IEC/EN 60601-1-1. Everybody who connects additional equipment to the signal
input connector or signal output connector configures a medical system, and is therefore responsible
that the system complies with the requirements of the valid version of the system standard IEC/EN
60601-1-1. If in doubt, contact our technical service department or your local CooperSurgical
representative.
DB9 Interface
RJ45 Interface
Pin Signal
Input/Output
1 +5V
Output
2 Rx
Input
3 Tx
Output
4 485EN
Input
5 0V
Ref.
6 TA
Output
7 TB
Output
8 RA
Input
9 RB
Input
Pin Signal
Input/Output
1 TD+
Output
2 TD-
Output
3 RD+
Input
4 Reserved
5 Reserved
6 RD-
Input
7 Reserved
8 Reserved