F9 Fetal & Maternal Monitor • User Manual
- 2 -
•
The protective categories against electric shock of the patient connections are:
1) Ultrasound (FHR1, FHR2) 2) External TOCO 3) Fetal Movement Mark (FM)
4) Fetal Stimulator (FS)
This symbol indicates that the protection against electric shock of this instrument is Type B.
IUP (Intrauterine Pressure)
This symbol indicates that the protection against electric shock of this instrument is Type BF.
DECG (Direct Electrocardiography)
This symbol indicates that the protection against electric shock of this instrument is Type CF.
The monitor described in this User Manual is not protected against:
a)
The effects of defibrillator shocks
b)
The effects of defibrillator discharge
c)
The effects of high frequency currents
d)
The interference of electrosurgery equipment
1.3 Ultrasound Safety Guide
Fetal Use
The F9 Monitor is designed for continuous fetal heart rate monitoring during pregnancy and labor.
Clinical interpretation of the fetal heart rate traces can diagnose fetal and/or maternal problems and
complications.
Instructions for Use in Minimizing Patient Exposure
The acoustic output of the F9 Monitor is internally controlled and can not be varied by the operator
in the course of the examination. The duration of exposure is, however, fully under the control of the
operator. Mastery of the examination techniques described in the User Manual will facilitate
obtaining the maximum amount of diagnostic information with the minimum amount of exposure.
Exercising clinical judgment in monitoring low risk patients will avoid unnecessary insonation.
Содержание F9 Express
Страница 1: ......
Страница 2: ......
Страница 89: ...F9 Fetal Maternal Monitor User Manual 83 NOTES...
Страница 90: ......
Страница 91: ......