53
Firmware upgrade
To upgrade the firmware file, the configuration file and the parameters file of the sterilizer,
follow these steps:
1.
Connect the sterilizer to the computer using the USB serial cable.
2.
Turn on the sterilizer from the
main switch 13
.
3.
Turn on the computer and install the SterilClaveVxx.exe program following the
instructions displayed on the screen.
4.
Start the SterilClaveVxx.exe program.
5.
Choose the data transfer rate: click on 115200 bps.
6.
Type the number of the COM port to which the sterilizer is connected.
7.
Click Open.
8.
Click the Debug tab.
9.
Click Identify. If the numbers displayed in the box are different from zero the
connection has been established correctly; otherwise, check the number of the
COM port and the cable since the connection is not working.
10.
Click the Download tab to download the firmware file, the configuration file and
the parameter file, IN THIS ORDER.
11.
Click Download Firmware File and select the firmware file (.s19) to start
downloading. When the download is completed, the message OPERATION
SUCCESSFUL will be displayed.
12.
Click Download Configuration File, select the configuration file (.cmxcnf) to start
downloading.
13.
Click Download Parameters File, select the parameters file (.cmxpars) to start
downloading.
Click Close to close the connection and click Exit to exit the program.
Firmware upgrade
Содержание Sterilclave 6 B Speedy
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Mod 6 B SPEEDY Contents...
Страница 7: ...7 Contents...
Страница 8: ...8 Contents...
Страница 57: ...57 Diagrams and wirings...
Страница 58: ...58 Wiring diagram 6 B SPEEDY Diagrams and wirings...
Страница 60: ...60 STEAM PEAK diagram EV2 ON EV4 ON Internal coil ON steam is injected in the chamber 09 Diagrams and wirings...