43
Automatic water inlet does not work correctly
If the
automatic water inlet 29
(optional) is not working correctly, proceed as follows:
•
Check that the automatic water inlet is actually available since not all models are
supplied with this device.
•
Check the External Charge setting (ON) in the Technician - Accessories Menu.
•
Check the Self-filling setting (ON) in the Main Settings Menu (see USE AND
MAINTENANCE MANUAL, chapter OPERATION).
•
For configuration
A
from external tank, check that there is water in the tank; if there
is water and the AUTOMATIC WATER INLET ALARM is on, check that the inlet pipe is
properly positioned and touches the bottom of the tank; check that the pipe is not
bent or twisted; check operation of the
automatic water inlet pump 46
;
•
In some cases, if the
automatic water inlet pump 46
runs dry for a long period of
time, performances may degrade and priming and filling the unit will not be
possible in the expected time. In this case, replace the pump.
•
For configuration
B
from water mains, check that the mains water tap is open.
If the tap is open and the AUTOMATIC WATER INLET ALARM is on, check for flow of
water downstream of the installed demineralization system. Check the inlet
solenoid valve
EV7
.
•
Check the fuse
F1
on the
power board 34.
This fuse
protects all loads. Replace it, if
necessary.
Printer does not print
Check that the
printer 26
is present and that Printer ON is selected in the Technician -
Accessories Menu, then proceed as follows:
•
Check that Printer setting in the Main Settings Menu allows use of the printer (see
USE AND MAINTENANCE MANUAL, chapter OPERATION).
•
Check that the internal flap of the printer containing the roll is closed.
•
Check that the paper roll is present and correctly fitted.
Note:
Never pull paper out as this may open the internal flap or tear the paper out of
position. To eject the entire printout, press FEED (in STAND BY mode) on the printer control
panel. The paper will be fed by a small amount each time the button is pressed.
•
Check
fuses 32
and fuse
F1
on the
power board 34.
Check the connection
between the board and the printer (see chapter Printer installation).
Additional checks and settings
Содержание Sterilclave 6 B Speedy
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Mod 6 B SPEEDY Contents...
Страница 7: ...7 Contents...
Страница 8: ...8 Contents...
Страница 57: ...57 Diagrams and wirings...
Страница 58: ...58 Wiring diagram 6 B SPEEDY Diagrams and wirings...
Страница 60: ...60 STEAM PEAK diagram EV2 ON EV4 ON Internal coil ON steam is injected in the chamber 09 Diagrams and wirings...