36
Generator insufficient vacuum alarm
This indicates that during the pre-vacuum phase in the
steam generator 47
, pressure did
not drop to the correct predetermined value.
•
Check the efficiency and cleanliness of the
door gasket 31
. Replace if damaged.
•
Check the efficiency and cleanliness of the gasket of the
steam generator 47
. Replace
if damaged.
•
Check operation of the
vacuum pump 41
.
•
Check the cleanliness and correct positioning of the
drain filter 20
.
•
Run a VACUUM TEST to check for pressure losses.
•
Check that ventilation air vents are clear and check efficiency of internal fans.
Generator warm up alarm
It informs about a problem in the warm up of the
steam generator 47
.
The sterilizer displays this alarm when the warm up of the steam generator 47 has not been
completed within the maximum time set by the corresponding parameter.
•
Check the efficiency and cleanliness of the gasket of the
steam generator 47
.
•
Check the
temperature probe PT2 39A
.
•
Check the efficiency and cleanliness of the internal coil of the
steam generator 47
.
•
Check the solenoid valve
EV4
.
Generator level alarm
This alarm indicates that the time has been exceeded for the correct level of water to be
reset in the
steam generator 47
.
The sterilizer displays this alarm when the water inlet pump 35 is supplied for a period of
time exceeding the time set in the corresponding parameter.
•
Check:
- that the
water inlet pump 35
and the solenoid valve
EV1
are correctly feeding water in
the
steam generator 47
.
- that the coil of the solenoid valve
EV1
is not burnt.
- the
water inlet pump 35
.
- the
level probes 43A
and
43B
of the
steam generator 47;
- the water filter;
-
the fuse
F1
on the
power board 34
.
Generator insufficient water level alarm
This alarm indicates that the
steam generator 47
does not contain enough water to
complete the cycle.
The sterilizer displays this alarm when the minimum level probe of the steam generator 47 is
open.
•
Fill the tank.
•
Check the efficiency and cleanliness of the
chamber gasket 31
. Replace if damaged.
If leakage of steam is found, adjust the
door 16
closing.
•
Check the efficiency and cleanliness of the
level probes 43A
and
43B
of the
steam
generator 47.
•
Check that the maximum load is not exceeded.
•
If the alarm is not resolved, check the drain solenoid valve EV2.
Alarms
Содержание Sterilclave 6 B Speedy
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Mod 6 B SPEEDY Contents...
Страница 7: ...7 Contents...
Страница 8: ...8 Contents...
Страница 57: ...57 Diagrams and wirings...
Страница 58: ...58 Wiring diagram 6 B SPEEDY Diagrams and wirings...
Страница 60: ...60 STEAM PEAK diagram EV2 ON EV4 ON Internal coil ON steam is injected in the chamber 09 Diagrams and wirings...