35
Manual stop alarm
This indicates that STOP was pressed during a cycle before the cycle was completed.
The sterilizer displays this alarm during operation and stops to inform the user that the load
MAY NOT BE STERILE.
•
This is not a failure but a cycle stop override caused by the operator. For this reason,
there are no operations to be performed but the load MAY NON BE STERILE. A new
sterilization cycle must be started.
Note:
Manual cycle stops should always be avoided unless absolutely necessary. If the
sterilizer has already performed the fractioned vacuum or pressurisation phases, the
material inside could be wet and the sterilizer may not start properly. Dry the material well
and change the bags.
Power missing alarm
This indicates that a power blackout occurred during a cycle.
The sterilizer displays this alarm when power is restored. The cycle and the phase when
the blackout occurred will appear on the display.
If the blackout occurred while an alarm was issued, the last alarm issued will be
displayed.
•
On the basis of the message on the display, you can infer if the sterilizer managed to
complete the sterilization cycle or not despite the power blackout.
A blackout may be caused by various reasons:
- the sterilizer was accidentally switched off by means of the
main switch 13
;
- mains blackout;
-
power cable 03
not plugged in correctly or damaged;
- blown
fuses 32
.
PT2 overtemperature alarm
The alarm indicates that temperature inside the
steam generator 47
has exceeded 170°C.
The sterilizer displays this alarm when the temperature of the steam generator 47 reaches
170°C.
•
Check that the wires of the
temperature probe PT2 39A
are neither disconnected, nor
worn or faulty.
•
Check that the load is positioned correctly on the supports.
PT2 temperature probe alarm
This is triggered when the
temperature probe PT2 39A
of the
steam generator 47
is
disconnected or faulty.
The sterilizer shows this alarm during operation when the temperature probe PT2 39A
detects a temperature of less than 3°C.
•
Check the connection of the
chamber temperature probe PT2
39A
to the
connector CN13.
•
Replace the probe if faulty.
Alarms
Содержание Sterilclave 6 B Speedy
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Mod 6 B SPEEDY Contents...
Страница 7: ...7 Contents...
Страница 8: ...8 Contents...
Страница 57: ...57 Diagrams and wirings...
Страница 58: ...58 Wiring diagram 6 B SPEEDY Diagrams and wirings...
Страница 60: ...60 STEAM PEAK diagram EV2 ON EV4 ON Internal coil ON steam is injected in the chamber 09 Diagrams and wirings...