45
•
Mount the
automatic water inlet pump 46
on its base, strapping it to the supports.
•
Fit the branches (
Y
) and all connecting pipes.
•
Fit the external filling and waste pipes; the end of the inlet pipe is cut in a particular
manner to prevent a suction effect when it rests on the bottom of the tank.
The waste pipe is provided with a quick coupling male attachment.
•
Test tightness.
•
In the Technician Menu, Accessories option (option 16), set External Charge ON;
check operation of the sterilizer. Close and restart the unit.
Automatic inlet
Содержание Sterilclave 6 B Speedy
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Mod 6 B SPEEDY Contents...
Страница 7: ...7 Contents...
Страница 8: ...8 Contents...
Страница 57: ...57 Diagrams and wirings...
Страница 58: ...58 Wiring diagram 6 B SPEEDY Diagrams and wirings...
Страница 60: ...60 STEAM PEAK diagram EV2 ON EV4 ON Internal coil ON steam is injected in the chamber 09 Diagrams and wirings...