Connect the calibration tool 4-pin cable to the connector CN11 of the
power cable 34.
Set the calibration tool switch to 0 V (position 2).
Wait for at least 10 seconds and press ENTER.
Now set the calibration tool switch to 5 V (position 1).
Wait for at least 10 seconds and press ENTER.
Press
Å
to go back to the Technician Menu without changing the calibration.