39
ADDITIONAL CHECKS AND SETTINGS
Display off or not clear
Proceed as follows if no writing appears on the display and/or if the display is not very
clear:
•
Check that power socket is powered, the
power cable 03
is correctly plugged in
and connected to the sterilizer.
•
Check the
fuses 32
and check that the
main switch 13
is pressed to position “I”.
•
Check the fuses on the
electronic power board 34
.
•
If the display is not clear from where you are standing, change reading position or
angle. If the display is clearer, adjust contrast by means of the trimmer
PT1
starting
from your preferred reading position (see chapter DIAGRAMS AND WIRING,
Electronic logic board).
Level probe check
The level probes of the
clean water tank 10
and of the
steam generator 47
are metal rods
fitted into the container to be monitored and insulated from the container itself. The probe
circuit is closed when liquid is present.
The presence of fouling or lime scales in the
clean water tank 10
may create electric
continuity between tank metal and level probe rod and consequently detect the
presence of liquid in the tank when the tank is empty.
If messages and alarms appear after emptying the
clean water tank 10
remove the
probes and clean them carefully as follows.
Follow the same procedure to check the level probes of the
steam generator 47.
Additional checks and settings
Содержание Sterilclave 6 B Speedy
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Mod 6 B SPEEDY Contents...
Страница 7: ...7 Contents...
Страница 8: ...8 Contents...
Страница 57: ...57 Diagrams and wirings...
Страница 58: ...58 Wiring diagram 6 B SPEEDY Diagrams and wirings...
Страница 60: ...60 STEAM PEAK diagram EV2 ON EV4 ON Internal coil ON steam is injected in the chamber 09 Diagrams and wirings...