To calibrate sensor PT2 or PT3, switch pole number 1 of the temperature selector to OFF
and switch pole number 3 of the temperature selector to ON, which corresponds to
a 1832 Ohm sample coil (220°C).
Wait for at least 10 seconds and press ENTER.
Press
Å
to go back to the Technician Menu without changing the calibration.