42
Vacuum pump servicing procedure
The
vacuum pump 41
can be serviced by cleaning it and replacing the gaskets which are
supplied in the specific kit.
•
Remove the pump and carefully check the
position of components.
•
Check that the membranes
are not cracked.
Clean them or replace them, if needed.
•
Check the
rubber valves. Clean them or replace
them, if needed. Make sure position and
orientation are correct when refitting.
Vacuum pump kit 2119:
Vacuum pump 2119:
Additional checks and settings