37
PT3 overtemperature alarm
The alarm indicates that temperature of the external coil
has exceeded 220°C.
The sterilizer displays this alarm when the temperature of the external coil has reached
220°C.
•
Check that the wires of the
temperature probe PT3 39B
are neither disconnected nor
worn and that the probe is not faulty.
PT3 temperature probe alarm
This is triggered when the
temperature probe PT3 39
of the external coil is disconnected or
faulty.
The sterilizer shows this alarm during operation when the sensor PT3 39B on the external coil
detects a temperature of less than 3°C.
•
Check the connection of the
temperature probe PT3
39B
to the connector CN14.
•
Replace the probe if faulty.
Events alarm
This alarm indicates that one of the timed cycles (sterilization, drying or maintenance in
the Vacuum Test) was not completed successfully.
•
Contact COMINOX.