25
MESSAGES
Messages contain information and warnings and usually require the user to perform a
simple correction or check.
Message/ Cause /
•
Solution
Scheduled maintenance
This message indicates the need for a scheduled maintenance on one or more
components.
This message will appear when the sterilizer is in STAND BY mode and the remaining cycles
or days of life of one or more components are expiring.
•
The sterilizer may however allow to run one or more cycles but the component
subject to scheduled maintenance must be replaced (see chapter MAINTENANCE).
Check water quality
This message appears only on models fitting an external water demineralization system
(optional) or if the water is automatically loaded from an external tank (optional); in this
case, the sterilizer monitors conductivity of loaded water.
The sterilizer will display this message in STAND BY mode and only when an automatic water
inlet 29 is installed (see USE AND MAINTENANCE MANUAL, chapter CONNECTIONS) and
Self-filling in the Main Settings Menu is ON. This indicates that the water conductivity has
exceeded the optimal level (15 µS/cm) but is still acceptable.
•
The sterilizer may allow to run one or more cycles but it is advisable to proceed as
described in section INSUFFICIENT WATER QUALITY because this alarm could stop
operation of the sterilizer.
Note: If the water conductivity increases, it means that the mineral salts content has
increased and water is therefore NO LONGER suitable.
Cooling pause
This message indicates that the temperature in
chamber
09
is too high to start or continue
a cycle (see USE AND MAINTENANCE MANUAL, chapter OPERATION).
The sterilizer will display this message in STAND BY mode or after starting a cycle and will
stop only during the pre-vacuum phase. This message normally appears when several
cycles are run one after the other and no time is allowed for the chamber to cool down.
•
In such conditions, simply let the chamber cool down. To speed up cooling, open
door
16
, otherwise cooling will take considerably longer. To open the door, press STOP,
cancel the MANUAL STOP ALARM and wait for a few seconds for the pressure inside
the chamber to equalise ambient pressure to allow the door to be opened.
•
With the door closed
the operating cycle will resume automatically when the allowed
temperature is reached.
•
If you opened the door
, start the cycle again after cooling down the sterilizer.
Messages
Содержание Sterilclave 6 B Speedy
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Mod 6 B SPEEDY Contents...
Страница 7: ...7 Contents...
Страница 8: ...8 Contents...
Страница 57: ...57 Diagrams and wirings...
Страница 58: ...58 Wiring diagram 6 B SPEEDY Diagrams and wirings...
Страница 60: ...60 STEAM PEAK diagram EV2 ON EV4 ON Internal coil ON steam is injected in the chamber 09 Diagrams and wirings...