26
8.
ZERO TRACKING
(1) From the mass display, press the CAL/MENU key several times.
When
"FUnC.SEL"
is displayed, press the
→
O/T
←
key.
(2) Press the CAL/MENU key several times until
"trC:**"
is displayed.
The ** indicates the current setting.
"on"
indicates zero tracking is
activated, and
"oF"
indicates it is released.
(3) Press the
→
O/T
←
key to change the
"trC:**"
setting.
(4) Each time CAL/MENU is pressed,
"trC-oF"
or
"trC-on"
is
displayed. The current setting is displayed with the stability symbol.
When the desired item is displayed, press the
→
O/T
←
key to
change the setting, or press the POWER/BRK to return to
"trC:**."
(5)
When
"trC:**"
is displayed, press and hold the POWER/BRK key
to return to the mass display.
9. STABILITY DETECTION BAND
(1) From the mass display, press the CAL/MENU several times. When
"FunC.SEL"
is displayed, press the
→
O/T
←
key.
(2) Press the CAL/MENU key several times. When
"bAnd:**"
is
displayed, press the
→
O/T
←
key. The ** indicates the current
setting.
(3) Each time CAL/MENU is pressed,
"b-1"
,
"b-5"
, and
"b-10"
are
displayed. The current setting is displayed with the stability symbol.
When the desired item is displayed, press the
→
O/T
←
key to
change the setting.
(4) Press the POWER/BRK key to return to
"bAnd:**."
10. REGISTRATION, RELEASE, AND SELECTION OF MEASUREMENT
UNITS
Registration of Measurement Units
(1) From the mass display, press the CAL/MENU several times. When
"FUnC.SEL"
is displayed, press the
→
O/T
←
key.
(2) Press the CAL/MENU key several times. When
"Unit.SEL"
is
displayed, press the
→
O/T
←
key.
(3) Each time the CAL/MENU key is pressed, a unit that can be
registered is displayed.
Currently registered units are displayed with the stability symbol.