32
(3) To change the setting, press the
→
O/T
←
key while
"AdiSP:**"
is
displayed.
(4) Each time the CAL/MENU key is pressed,
"Ad-oF"
or
"Ad-on"
is
displayed. The stability symbol illuminates when the current setting
is displayed.
(5) Press the
→
O/T
←
key to change the setting, or press the
POWER/BRK key to return to
"AdiSP:**"
without changing the
settings.
(6) Press and hold the POWER/BRK key to return to the mass display.
13. ADD LOAD WEIGHING
This function displays (or prints) the individual weights and summarizes the
total weight of a series of additions to the sample pan. The
Addon:**
function must be reset for each series of Add Load Weighing.
Auto PRINT function must be set to ON to use this function. The balance
will return to the zero display automatically when the printer is attached. If a
printer is not attached, press the
→
O/T
←
key to manually return the display
to zero after each addition. Press the POWER/BRK to send the total weight
to the printer and return to normal weighing mode.
With printer attached:
Sample weight is sent to the printer upon stabilization, followed by
automatically returning to the zero display. As additional material is added
to the sample, the additional weight is sent to the printer upon stabilization
and the balance returns to the zero display. Press the POWER/BRK key to
send the total weight to printer and return to normal weighing mode.
Without printer attached:
Auto PRINT function must be set ON to use this function. Press the
→
O/T
←
key after each addition is displayed. Press the POWER/BRK key
to display total weight and return to normal weighing mode.
Reset the
Addon:**
function for each series of Add Load Weighing.
Procedure
(1) Press the CAL/MENU key several times from the mass display.
When
"FUnC.SEL"
is displayed, press the
→
O/T
←
key.
(2) Press the CAL/MENU several times to display
"Addon:**."
The ** indicate the current setting for the Add Load Weighing
function.
"on"
indicates that Add Load Weighing is active and
"oF"
indicates that it is released.