15. EXTERNAL WEIGHT VALUE FOR SPAN CALIBRATION
Procedure
(1) From the mass display, press the CAL/MENU key repeatedly until
"SEttinG"
is displayed, press the
→
O/T
←
key.
(2) Press the CAL/MENU key several times. When
"CAL SEt"
is
displayed, press the
→
O/T
←
key. The [MENU] and # symbols are
displayed to indicate numeric value entry. Depending on the model,
50.0000, 100.0000, 200.0000 or 300.0000 is displayed with the
leftmost digit blinking.
(3) Change the blinking digit by pressing the UNIT key. Each time the
UNIT key is pressed, the value changes from 1
→
2
→
• • •
→
8
→
9
→
0
→
1
→
• • •. Press the PRINT key to confirm that value
and move the blinking cursor to the next digit.
(4) Press the
→
O/T
←
key to confirm the numeric value of the external
weight for span calibration or press the POWER/BRK key to return
to
"CAL SEt"
without changing the value.
(5) When
"CAL SEt"
is displayed, press and hold the POWER/BRK
key to return to mass display.
16. EXTERNAL WEIGHT VALUE FOR BUILT-IN WEIGHT
CALIBRATION (
CAW/CAX series only)
Procedure
(1) From the mass display, press the CAL/MENU key repeatedly until
"SEttinG"
appears, press the
→
O/T
←
key.
(2) Press the CAL/MENU key several times. Whe n
"PCAL.SEt"
is
displayed, press the
→
O/T
←
key. The [MENU] and # symbols are
displayed to indicate numeric value entry. Depending on the model,
50.0000, 100.0000, 200.0000 or 300.0000 is displayed with the
leftmost digit blinking.
(3) Change the blinking digit by pressing the UNIT key. Each time the
UNIT key is pressed, the value changes from 1
→
2
→
• • •
→
8
→
9
→
0
→
1
→
• • •. Press the PRINT key to confirm that value
and move the blinking cursor to the next digit.
(4) Press the
→
O/T
←
key to confirm the numeric value of the external
weight for span calibration or press the POWER/BRK key to return
to
"PCAL.SEt"
without changing the value.
(5) When
"PCAL.SEt"
is displayed, press and hold the POWER/BRK
key to return to mass display.