27. PRINTING THE DATE & TIME
(CAW/CAX series only)
It is possible to set the balance to send the date and time to a peripheral
device such as a printer. This records the date and time of the internal clock
on the span calibration record.
Procedure
(1) Press the CAL/MENU key several times from the mass display.
When
"SEttinG"
is displayed, press the
→
O/T
←
key.
(2) Press the CAL/MENU several times to display
"Prtdt:**."
The **
indicate the current setting for date and time printing.
"on"
indicates
that the function is active and
"oF"
indicates that the function is
released.
(3) To change the setting, press the
→
O/T
←
key while
"Prtdt:**"
is
displayed.
(4) Each time the CAL/MENU key is pressed the display switches
between
"Prtdt-oF"
and
"Prtdt-on"
.
(5) Press the
→
O/T
←
key to change the setting or press the
POWER/BRK key to return to
"Prtdt:**"
without changing the
setting.
(6) Press and hold the POWER/BRK key while
"Prtdt:**"
is displayed,
to return to mass display.
28. PERFORMANCE CHECK
Performance checks should be conducted in an environment where the room
temperature is stable at around 25°C. Use these performance checks to
determine whether the balance is operating properly.
Preparation
•
Verify that the balance is warmed-up.
F
See Section
3.
“WARM-UP.”
•
Set the following measuring conditions:
Stnd
(Standard mode)
b-1
(Stability detection band 1 count)
trC-oF
(Zero tracking OFF)